Interventional Room Pendant Selection Guide: Electric Pendants, CO2 Interfaces and AGSS Planning
WEYUAN Medical / News and Knowledge
Interventional room pendant selection is not only a question of catalogue parameters. The real test is whether doctors and nurses can work smoothly during long procedures, with gas, power and equipment exactly where the workflow needs them.
For new catheterization laboratories, interventional suites and renovation projects, the pendant system has a direct effect on procedure continuity. The key decisions are often missed during procurement: electric movement, CO2 terminals, AGSS planning and the number of pendants required for the room type.
Electric pendant or mechanical pendant
Interventional procedures require frequent pendant adjustment. A doctor may be holding a catheter, guidewire or sheath while watching the imaging screen. Any movement that forces the circulating nurse to leave the sterile workflow interrupts the procedure rhythm.
With a mechanical pendant, the nurse usually needs to step away, push or pull the arm, move the box, lock it again and then return to the case. In a complex procedure lasting six to eight hours, this may happen many times. Even a one minute interruption repeated across the day becomes real lost operating time.
An electric pendant solves this more directly. Height adjustment is controlled by button, so the team can reposition the terminal box with less disruption. Electromagnetic braking also improves positioning accuracy. Mechanical or pneumatic locking often allows positioning tolerance around +/-50 mm, while electric systems can normally control movement within a much tighter range, depending on load and installation conditions.
Why the CO2 terminal matters
The CO2 terminal is one of the clearest differences between an interventional pendant and a standard surgical pendant. In structural heart procedures and electrophysiology work, CO2 may be used for bubble imaging. Because CO2 is radiolucent under X-ray, it can help outline internal structures and support catheter positioning.
CO2 is used instead of air or oxygen because it dissolves rapidly in blood. Air introduced into coronary or cerebral circulation can create serious embolic risk. For interventional teams, CO2 access is therefore not a decorative option. It can be a clinical requirement for specific procedure types.
A surgical pendant generally does not require CO2 because ordinary open surgical workflows do not use this form of gas imaging. That is why CO2 can be treated as an identity marker for an interventional pendant.
AGSS should be planned early
Whether an anesthetic gas scavenging system, or AGSS, is required depends on current and future procedure scope. In local anesthesia rooms, AGSS may not be used every day. Even then, reserving the pipeline and terminal position during construction is usually the safer economic decision.
If the room later adds pediatric interventions, complex TAVR procedures or sedation support, the anesthesia machine or sedation workstation may need to connect near the pendant. Retrofitting AGSS after the ceiling, gas system and laminar-flow environment are complete costs far more than reserving the interface during construction.
When a sedation workstation is mounted on the operator-side pendant, AGSS changes from a reserve item to a required configuration if waste anesthetic gas is produced. If an anesthesia machine is placed under the pendant during complex procedures, AGSS is also a necessary safety interface.
How many pendants does the room need?
A standard interventional room should normally plan at least two electric pendants, one on the operator side and one on the assistant or ultrasound side. The operator-side pendant may carry DSA displays, injector equipment, electrophysiology recording equipment and a sedation workstation. The assistant-side pendant may support ultrasound, backup monitoring and infusion equipment.
If the room is a true hybrid operating room with open surgical capability, a third pendant may be required. The anesthesia pendant supports airway and anesthesia monitoring. The surgical pendant supports electrosurgical equipment, suction and surgical tools. The interventional pendant supports imaging and catheter-based equipment.
Typical gas and power planning
| Configuration item | Interventional pendant | Surgical pendant | Anesthesia pendant |
|---|---|---|---|
| O2 | 2 outlets, including nasal cannula and anesthesia backup | 2 outlets | 2 outlets |
| AIR | 1 outlet for pneumatic equipment | 1 outlet for drill or saw systems | 1 outlet |
| CO2 | 1 required terminal | Usually not required | Reserved when needed |
| VAC | 1 to 2 high-vacuum outlets | 1 to 2 outlets | 1 independent suction outlet |
| AGSS | Required if sedation or anesthesia machine connects nearby | Usually not required | Required |
| Power sockets | 10 to 12 | 8 to 10 | About 8 |
The practical conclusion is simple: decide the pendant type by room function, not by the lowest equipment price. A standard interventional room can often work with two electric pendants. A true hybrid room should be planned with three functional pendant zones.